Scilogex DM0636 User Manual page 22

Multi-purpose centrifuge
Table of Contents

Advertisement

USER MANUAL
ID
Rotor type
code
1
AS30-2
2
AS60-2
3
AS18-5V
4
AS12-PCR8
5
AS6-50V
6
A30-15
7
A8-50
8
SE4-100A
S2-MP
9
(dimension
mm
*
:15mlV means 15ml conical tube, as well as 5mlV and 50mlV.
11.1.3 Notice
(1) The centrifuge rotor can separate sample which density lower than 2.0g/ml, if the samples density is
over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed× (2.0(g/ml)/Sample density (g/ml))
(2) To prevent corrosion, remove the rotor from rotor chamber if do not use for a long time, then detach the
rotor lid and upside the rotor down to dry the tube holes.
Tube/bottle
2/1.5ml ×30
0.2ml ×30
0.5ml×30
2/1.5ml ×60
0.2ml×60
0.5ml×60
5mlV×18
8-PCR ×12
0.2ml×96
50mlV×6
15mlV×30
15mlV×16
50mlV×8
×4
100ml
×4
85ml
A85P100
×4
50mlV
A50VP100
×8
15mlV
A15VP100
×8
3~10ml
A10P100
MTP
(128×85.6×15)
Cell culture
(128×85.6×21)
)
DWP
(128×85.6×45)
Kits
(128×85.6×60)
Table 11.1 Rotors and adapters
Max.
Max.centrifugal radius
Adapter
speed
(rpm)
6000
A02P2
6000
A05P2
6000
6000
A02P2
6000
A05P2
6000
6000
6000
6000
6000
4500
5000
4000
4000
4000
4000
4000
4000
4000
4000
4000
Max. RCF
Rcf (× g)
r
(cm)
max
10
8.5
9
10
8.5
9
10
10
10
10.7
R1=14.2
R2=12.2
12.4
15.9
15.9
15.9
15.1
14.7
12.1
12.1
12.1
12.1
1/2
4020
3415
3618
4020
3415
3618
4020
4020
4020
4306
3210
2760
3460
2840
2840
2840
2700
2630
2160
2160
2160
2160