Fluke 5080A Service Manual page 36

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

5080A
Service Manual
0.0000 Ω
1.0000 Ω
1.9000 Ω
10.000 Ω
19.000 Ω
100.000 Ω
190.000 Ω
1.00000 kΩ
1.90000 kΩ
10.0000 kΩ
19.0000 kΩ
100.000 kΩ
190.000 kΩ
1.00000 MΩ
1.90000 MΩ
10.000 MΩ
19.000 MΩ
100.00 MΩ
190.00 MΩ
33 mV
33 mV
33 mV
33 mV
33 mV
33 mV
330 mV
330 mV
330 mV
3.3 V
3.3 V
26
Table 7. Verification Tests for 4-Wire Resistance
Range
Table 8. Verification Tests for AC Voltage (Normal)
Range
Output
10.00 mV
10.00 mV
10.00 mV
32.90 mV
32.90 mV
32.90 mV
329.00 mV
329.00 mV
329.00 mV
3.2900 V
3.2900 V
Output
0.0000 Ω
1.0000 Ω
1.9000 Ω
10.000 Ω
19.000 Ω
100.000 Ω
190.000 Ω
1.00000 kΩ
1.90000 kΩ
10.0000 kΩ
19.0000 kΩ
100.000 kΩ
190.000 kΩ
1.00000 MΩ
1.90000 MΩ
10.000 MΩ
19.000 MΩ
100.00 MΩ
190.00 MΩ
Frequency
45 Hz
100 Hz
1 kHz
45 Hz
100 Hz
1 kHz
45 Hz
100 Hz
1 kHz
45 Hz
100 Hz
Tolerance
±0.01000 Ω
±0.00989 Ω
±0.00931 Ω
±0.014 Ω
±0.0171 Ω
±0.035 Ω
±0.0665 Ω
±0.00022 kΩ
±0.000418 kΩ
±0.0022 kΩ
±0.00494 kΩ
±0.035 kΩ
±0.0741 kΩ
±0.00035 MΩ
±0.000665 MΩ
±0.0092 MΩ
±0.0266 MΩ
±0.049 MΩ
±1.876 MΩ
Lower Limit
Upper Limit
9.909 mV
10.091 mV
9.908 mV
10.092 mV
9.908 mV
10.092 mV
32.738 V
30.062 V
32.735 V
33.065 V
32.735 V
33.065 V
328.512 V
329.488 V
328.479 V
329.521 V
328.479 V
329.521 V
3.28686 V
3.29314V
3.28653 V
3.29347 V