Fluke 5080A Service Manual page 51

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

Table 26. Leveled Sine Wave Function Verification: Flatness (cont.)
Nominal Value
Frequency
(p-p)
400.0 mV
1 MHz
400.0 mV
1 MHz
400.0 mV
2 MHz
400.0 mV
5 MHz
400.0 mV
10 MHz
400.0 mV
20 MHz
400.0 mV
50 MHz
400.0 mV
100 MHz
400.0 mV
125 MHz
400.0 mV
160 MHz
400.0 mV
200 MHz
1.3 V
500 kHz
1.3 V
1 MHz
1.3 V
1 MHz
1.3 V
2 MHz
1.3 V
5 MHz
1.3 V
10 MHz
1.3 V
20 MHz
1.3 V
50 MHz
1.3 V
100 MHz
1.3 V
125 MHz
1.3 V
160 MHz
1.3 V
200 MHz
5.5 V
500 kHz
5.5 V
1 MHz
5.5 V
1 MHz
5.5 V
2 MHz
5.5 V
5 MHz
5.5 V
10 MHz
5.5 V
20 MHz
5.5 V
50 MHz
5.5 V
100 MHz
Measured Value (p-p)
1-Year Spec.
Deviation (mV)
(mV)
6.10
6.10
6.10
6.10
6.10
6.10
6.10
6.10
8.10
8.10
8.10
19.60
19.60
19.60
19.60
19.60
19.60
19.60
19.60
19.60
26.10
26.10
26.10
82.5
82.5
82.5
82.5
82.5
82.5
82.5
82.5
82.5
Calibrator
Performance Tests
41