5080A
Service Manual
6.9800 MΩ
10.000 MΩ
10.240 MΩ
20.980 MΩ
39.190 MΩ
76.550 MΩ
100.00 MΩ
138.60 MΩ
148.90 MΩ
289.60 MΩ
559.60 MΩ
6. Connect the Product directly to the megohmmeter as shown in Figure 9.
When you use the 100 G
megohmmeters, the leads must be swapped between the HI and LO
ohms resistance output on the Product. The ground must be turned
on (Gnd On) when you swap HI and LO leads positions in the high
ohms resistance function. For example, to make a correct
measurement with the Quadtech 1865 Megohmmeter, connect the
HI terminal on the megohmmeter to the LO terminal on the Product
and connect the LO terminal on the megohmmeter to the HI terminal
on the Product. Turn the ground on and make the measurement.
32
Table 16. Megohm Option HVR Verification Point (cont.)
Range
Make sure you reverse the polarity.
Megohmmeter
Figure 9. High Resistance Verification for 1 G Ω and higher
Output
6.9800 MΩ
10.000 MΩ
10.240 MΩ
20.980 MΩ
39.190 MΩ
76.550 MΩ
100.00 MΩ
138.60 MΩ
148.90 MΩ
289.60 MΩ
559.60 MΩ
Note
Ω
value on the Product for some
Tolerance
±0.02094 MΩ
±0.0500 MΩ
±0.0512 MΩ
±0.1049 MΩ
±0.196 MΩ
±0.3827 MΩ
±0.500 MΩ
±0.693 MΩ
±0.744 MΩ
±1.448 MΩ
±2.798 MΩ
Fluke 5080A
gru003.eps