7.2.3.
The current measurement is forced to a preset value and all subsequent measurements are taken from this preset value. Preset can be applied
either from the WHT buttons or from the controller. To use the WHT button, the button state must be first set up – see section below on Button
allocation. Preset can only be cleared from the controller.
7.2.4.
The measurement from the WHT may be scaled by a multiplication factor, The scaling can only be set or cleared from the controller software
7.2.5.
The WHT can be set up so that the most positive measurement is displayed. This can be set up only from the controller. The MAX (PEAK+) can only
be reset from the controller
7.2.6.
The WHT can be set up so that the most negative measurement is displayed. This can be set up only from the controller. The MIN (PEAK-) can only
be reset from the controller
7.2.7.
The WHT can be set up so that the difference between MAX and MIN is transmitted. This can be set up only from the controller. The DIFF can only
be reset from the controller
All of the above modes are either continuous transmission or on a button press.