Setup Menus; Figure 15 Dcc Menu - Michell Instruments Optidew User Manual

Chilled mirror hygrometer
Hide thumbs Also See for Optidew:
Table of Contents

Advertisement

OPERATION
3.2

Setup Menus

DCC
Type
Setpoint
Mode
Interval
Period
Output hold
FAST
FAST SetP
20
Figure 15
DCC heating temperature can either be relative to last measured
dew point or absolute temperature. Actual temperature or Δ is
defined by 'Setpoint'.
Available input: Relative, Absolute
Mirror heating temperature during DCC, either absolute or
relative to last measured dew point. See 'Type' option below.
Available input: 1 to 120°C
DCCs can either be triggered automatically at every Interval, or
they can be manually triggered only.
Available input: Manual, Auto
Time between automatic DCCs
Input format: hh:mm
Limits: 01:00 to 99:00
Duration of a DCC
Input format: hh:mm
Limits: 00:01 to 00:59
Minimum time to hold analog outputs after finishing a DCC
Input format: hh:mm
Limits: 00:04 to 00:59
Turns frost assurance on or off. See section 4.4.3 for further
information
Available input: On, Off
Passing this mirror temperature will trigger the frost assurance
function without a DCC
Available input: -28 to -2°C
Optidew User's Manual
DCC menu
97551 Issue 1, February 2018