Instrutech The Hornet IGM402 User Manual page 72

Hot cathode ionization vacuum gauge with dual convection
Table of Contents

Advertisement

Instruction Manual
COMMAND
BRIEF DESCRIPTION
READ EMISSION
Read emission current
CURRENT STATUS
setting
SET EMISSION
Choose emission current of
CURRENT
4 mA or 100 μA
SET FILAMENT
Choose Filament 1 or 2
SET TRIP POINT for
Set the 'turns on below'
I
RELAY
(IG)
(+) pressure point for RLY
and set the 'turns off above'
(-) pressure point for RLY
as indicated by the pressure
display for the ion gauge
(IG)
SET TRIP POINT for
RLY A
Set the 'turns on below' (+)
pressure point for RLY A and
set the 'turns off above' (-)
pressure point for RLY A
Hint: Set the 'turns off
above' (-) setpoint pressure
first.
NOTE: RLY A/B may be
assigned to either CG1 or
CG2 pressure readings.
SET TRIP POINT
for RLY B
Set the 'turns on below' (+)
setpoint pressure for RLY B
and set the 'turns off above'
(-) setpoint pressure for RLY
B
Hint: Set the 'turns off
above' (-) setpoint pressure
first
READ TRIP POINT
Read the 'turns on below'
I
for RLY
(IG)
(+) pressure point for relay
and read the
'turns off above' (-) pressure
point for relay
READ TRIP POINT
Read the 'turns on below'
for RLY A
(+) pressure point for relay
A and read the
'turns off above' (-) pressure
point for relay A
InstruTech
COMMAND SYNTAX
#xxSES<CR>
(e.g., #01SES<CR>)
#01SEy<CR>
(e.g., #01SE1<CR>)
for y=1=4 mA; y=0=100μA
#xxSFy<CR>
(e.g., #01SF1<CR>)
y=1 for F1; y=2 for F2
#xxSLzy.yyEzpp<CR>
I
(e.g., #01SL+4.00E-06<CR>)
(e.g., #01SL-5.50E-06<CR>)
I
For the 'z' preceding the 'y.yy'
value, a '+' = 'turns ON below'
and a '-' = 'turns OFF above'
the setpoint value for RLY
#xxSLAzy.yyEzpp<CR>
(e.g., #01SLA+4.00E+02<CR>)
(e.g., #01SLA-5.00E+02<CR>)
Turn the relay OFF (-) above
500 Torr
#xxSLBzy.yyEzyy<CR>
(e.g., #01SLB+4.00E+02<CR>)
Turn the relay ON (+) below
400 Torr
(e.g., #01SLB-5.00E+02<CR>)
Turn the relay OFF (-) above
500 Torr
#xxRLz<CR>
I
(e.g., #01RL+<CR>)
(e.g., #01RL-<CR>)
I
#xxRLAz<CR>
(e.g., #01RLA+<CR>)
(e.g., #01RLA-<CR>)
IGM402 Hornet
RESPONSE
*xx_0.1MA_EM<CR>
*xx_4.0MA_EM<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
I
*xx_PROGM_OK<CR>
If the 'turns off above' (-)
pressure setpoint is less than
the value of the 'turns on
below' (+) setpoint, a syntax
error of ?01 SYNTX ER<CR> will
result.
*xx_PROGM_OK<CR>
If the 'turns off above' (-)
pressure setpoint is less than
the value of the 'turns on
below' (+) setpoint, a syntax
error of ?01 SYNTX ER<CR> will
result.
*xx_y.yyEzyy<CR>
(e.g., *01+7.60E+02<CR>
(e.g., *01-7.60E+02<CR>)
*xx_y.yyEzyy<CR>
(e.g., *01+5.60E+02<CR>)
(e.g., *01-7.60E+02<CR>)
Page 70