RS485 Command Protocol Summary - BINARY Protocol with CRC8

8.2
COMMAND
BRIEF DESCRIPTION
READ ALL
Read the current pressure for
PRESSURES
IG, CG1, and CG2. The
returned values are in
floating point.
Message Length: 17 bytes
READ CG
Read the current pressure for
PRESSURES
CG1 and CG2. The returned
values are in floating point.
Message Length: 13 bytes
READ IG PRESSURE
Read the current pressure for
ONLY
the IG. The returned value is
in floating point.
Message Length: 9 bytes
READ CG1
Read the current pressure for
PRESSURE ONLY
the CG1. The returned value
is in floating point.
Message Length: 9 bytes
READ CG2
Read the current pressure for
PRESSURE ONLY
the CG2. The returned value
is in floating point.
Message Length: 9 bytes
READ ION GAUGE
Determine if the ion gauge is
STATUS
on or off.
Message Length: 5 bytes
TURN ON THE ION
Turn ON the ion gauge
GAUGE
module.
Message Length: 5 bytes
InstruTech
COMMAND SYNTAX
!<01><00><xx><xx><xx><xx>
<xx><xx><xx><xx><xx>
<xx><xx><xx><xx><CRC>
<cmd>: 0x00
<xx>: represents the floating point
value place holder.
!<01><01><xx><xx><xx><xx>
<xx><xx><xx><xx><xx>
<CRC>
<cmd>: 0x01
<xx>: represents the floating point
value place holder.
!<01><02><xx><xx><xx><xx>
<xx><CRC>
<cmd>: 0x02
<xx>: represents the floating point
value place holder.
!<01><03><xx><xx><xx><xx>
<xx><CRC>
<cmd>: 0x03
<xx>: represents the floating point
value place holder.
!<01><04><xx><xx><xx><xx>
<xx><CRC>
<cmd>: 0x04
<xx>: represents the floating point
value place holder.
!<01><15><xx><CRC>
<cmd>: 0x15
<xx>: represents the byte value place
holder.
!<01><05><xx><CRC>
<cmd>: 0x05
<xx>: represents the byte value place
holder.
IGM402 Hornet
RESPONSE
*<01><00><uu><rr><rr>
<rr><rr><yy><yy><yy>
<yy><zz><zz><zz><zz>
<CRC>
<uu>: pressure units
<rr>: IG float value
<yy>: CG1 float value
<zz>: CG2 float value
*<01><01><uu><yy><yy>
<yy><yy><zz><zz><zz>
<zz><CRC>
<uu>: pressure units
<yy>: CG1 float value
<zz>: CG2 float value
*<01><02><uu><yy><yy>
<yy><yy><CRC>
<yy>: represents the four byte
floating point value.
<uu>: pressure units
*<01><03><uu><yy><yy>
<yy><yy><CRC>
<yy>: represents the four byte
floating point value.
<uu>: pressure units
*<01><04><uu><yy><yy>
<yy><yy><CRC>
<yy>: represents the four byte
floating point value.
<uu>: pressure units
ON:
*<01><15><01><CRC>
OFF:
*<01><15><00><CRC>
ON:
*<01><05><01><CRC>