Instruction Manual
COMMAND
BRIEF DESCRIPTION
SET UNIT ADDRESS
Set the address of the
unit.
Message Length: 5 bytes
SET UNIT ADDRESS
Set the address offset of the
OFFSET
unit.
Message Length: 5 bytes
I
READ RELAY
HIGH
Get the pressure when
TRIP POINT
Relay
Message Length: 8 bytes
SET RELAY I HIGH
Set the pressure when
TRIP POINT
Relay
Message Length: 8 bytes
I
READ RELAY
LOW
Get the pressure when
TRIP POINT
Relay
Message Length: 8 bytes
I
SET RELAY
LOW
Set the pressure when
TRIP POINT
Relay
Message Length: 8 bytes
READ RELAY A
Get the pressure when
HIGH TRIP POINT
Relay A will turn off.
Message Length: 8 bytes
(1)
Commands marked with a (1) under the "BRIEF DESCRIPTION" column will not take effect until after RESET
command is sent or power is cycled.
InstruTech
(1)
(1)
I
will turn off.
I
will turn off.
I
will turn on.
I
will turn on
COMMAND SYNTAX
!<01><38><xx><CRC>
<cmd>: 0x38
<xx>: represents the address offset
value.
!<01><39><xx><CRC>
<cmd>: 0x39
<xx>: represents the address offset
value.
!<01><26><xx><xx><xx><xx><CRC>
<cmd>: 0x26
<xx>: represents the four byte
floating point value place holder.
!<01><0F><xx><xx><xx><xx><CRC>
<cmd>: 0x0F
<xx>: represents the four byte
floating point value.
!<01><27><xx><xx><xx><xx><CRC>
<cmd>: 0x27
<xx>: represents the four byte
floating point value place holder.
!<01><10><xx><xx><xx><xx><CRC>
<cmd>: 0x10
<xx>: represents the four byte
floating point value.
!<01><28><xx><xx><xx><xx><CRC>
<cmd>: 0x28
<xx>: represents the four byte
floating point value place holder.
IGM402 Hornet
RESPONSE
*<01><38><xx><CRC>
*<01><39><xx><CRC>
<xx>: represents the address
offset value.
*<01><26><yy><yy><yy>
<yy><CRC>
<yy>: represents the four byte
floating point value.
*<01><0F><xx><xx><xx>
<xx><CRC>
*<01><27><yy><yy><yy>
<yy><CRC>
<yy>: represents the four byte
floating point value.
*<01><10><xx><xx><xx>
<xx><CRC>
*<01><28><yy><yy><yy>
<yy><CRC>
<yy>: represents the four byte
floating point value.
Page 62