3.3.2.3 Analog Output And Setpoint Relays For Cg1 And Cg2 - Instrutech The Hornet IGM402 User Manual

Hot cathode ionization vacuum gauge with dual convection
Table of Contents

Advertisement

Instruction Manual

3.3.2.3 Analog output and setpoint relays for CG1 and CG2

When using analog outputs and set-point relays for CG1 and CG2, use the green terminal blocks as described
below:
Connection
A +
A -
NO
COM
NC
CG1 Output Signal
and Relay A Contacts
Setpoint Relay - NC
Setpoint Relay - COM
Setpoint Relay - NO
Analog Output Ground A-
Analog Output A+
Connect to CG1 Cable
See
section 5.8.4.
– ASSIGN RLY A and ASSIGN RLY B for details on programming the CG device assigned to
control RLY A and RLY B.
InstruTech
IGM402 CG1 and CG2 convection gauge I/O connectors
Description
Analog outputs for convection gauges CG1 or CG2:
log-linear 1 to 8 Vdc, 1 V/decade or Non-Linear S-Curve 0.375 to 5.659 Vdc
Analog signal ground. Use only as analog output ground.
Setpoint relays for CG1 or CG2 - Normally Open contact
Setpoint relay - Common contact (program screen for relays = RLY A / RLY B)
Setpoint relays for CG1 or CG2 - Normally Closed contact
IGM402 I/O Terminal Blocks for CG1, CG2, RELAY A and RELAY B
IGM402 Hornet
CG2 Output Signal
and Relay B Contacts
Analog Output A+
Analog Output Ground A-
Setpoint Relay - NO
Setpoint Relay - COM
Setpoint Relay - NC
Connect to CG2 Cable
Page 17