Instrutech The Hornet IGM402 User Manual page 69

Hot cathode ionization vacuum gauge with dual convection
Table of Contents

Advertisement

Instruction Manual
COMMAND
BRIEF DESCRIPTION
READ CG2 ANALOG
Get the type of analog
OUTPUT TYPE
output setup for CG2
Message Length: 5 bytes
SET CG2 ANALOG
Set the type of analog
OUTPUT TYPE
output setup for CG2
Message Length: 5 bytes
µ
Read the pressure where
READ 100
A TURN
CG1 will turn on IG
ON PRESSURE
Message Length: 8 bytes
µ
Set the pressure where
SET 100
A TURN
CG1 will turn on IG
ON PRESSURE
Message Length: 8 bytes
It is recommended that if you are not familiar with the use of the BINARY protocol described in this section,
please consider using the ASCII type protocol described in the following sections 8.3 and 8.4. The BINARY
protocol described above is a sophisticated serial communication protocol used only when control system
designs require the use of cyclic redundancy checking (transmission error checking) of the commands and
responses sent and received between master and slave devices. Using the BINARY protocol requires special
programming knowledge and capability.
InstruTech
COMMAND SYNTAX
!<01><37><xx><CRC>
<cmd>: 0x37
<xx>: represents the byte value
place holder.
!<01><36><xx><CRC>
<cmd>: 0x36
<xx>: represents the byte value as
follows:
00:
Non-Linear
01:
Log-Linear
!<01><43><xx><xx><xx><xx>
<CRC>
<cmd>: 0x43
<xx> Represents a byte value place
holder
!<01><44><xx><xx><xx><xx>
<CRC>
<cmd>: 0x44
<xx> Represents a byte value place
holder
IGM402 Hornet
RESPONSE
*<01><37><xx><CRC>
<yy>: represents the byte
value as follows:
00:
Non-Linear
01:
Log-Linear
*<01><36><xx><CRC>
*<01><43><yy><yy><yy>
<yy><CRC>
<yy> Represents a byte value
place holder for a floating
point number
*<01><44><yy><yy><yy>
<yy><CRC>
<yy> Represents a byte value
place holder for a floating
point number
Page 67