wtw LF 170 Operating Instructions Manual page 26

Conductivity monitor
Table of Contents

Advertisement

Operation
Setting
CR1
(continued)
92
Selection/Values
Explanation
(values in bold typeface = values on delivery)
The following settings are possible:
, Measuring range 0.000 ... 1.999 µS/cm:
Upper final value:
Lower final value:
(fixed range)
, Measuring range 0,00 ... 19,99 µS/cm:
Upper final value:
Lower final value:
Minimum spacing:
(step size 1 µS/cm)
, Measuring range 0.0 ... 199.9 µS/cm:
Upper final value:
Lower final value:
Minimum spacing:
(step size 10 µS/cm)
, Measuring range 0 ... 1999 µS/cm:
Upper final value:
Lower final value:
Minimum spacing:
(step size 100 µS/cm)
, Measuring range 0.00 ... 19.99 mS/cm:
Upper final value:
Lower final value:
Minimum spacing:
(step size 1 mS/cm)
(continues on the next page)
LF 170
0,000 µS/cm
1.999 µS/cm
2.00 ... 19.99 µS/cm
0.00 ... 15.00 µS/cm
2 µS/cm
20.0 ... 199.9 µS/cm
0.0 ... 150.0 µS/cm
20 µS/cm
500 ... 1999 µS/cm
0 ... 1500 µS/cm
500 µS/cm
5.00 ... 19.99 mS/cm
0.00 ... 14.99 mS/cm
5 mS/cm
ba54104e09
11/2010