Keithley 6517 Service Manual page 87

Electrometer
Table of Contents

Advertisement

Troubleshooting
Table 5-5
Analog board checks (cont.)
Step Item/component
40
2V ANALOG OUTPUT
41
Front panel controls
42
INPUT jack
43
PREAMP OUT
44
U215 pin 6
45
2V ANALOG OUTPUT
46
Front panel controls
47
U308 pin 4
48
V SOURCE HI
49
Front panel controls
50
U308 pin 4
51
V SOURCE HI
52
Q311 drain
53
CR302 cathode
54
VR311 anode
55
CR308 cathode
5-14
Required condition
+2V
Coulombs function, 20nC range, zero check off.
Apply 20nC charge.
+20V
-10V
+2V
Set voltage source to 100V, operate on.
4.88V
100V
Set voltage source to 100V, operate on.
4.81V
1000V
+1200V
+120V
-1200V
-120V
Remarks
Referenced to COMMON.
Initial test conditions.
Input test charge.
Referenced to COMMON.
Referenced to COMMON.
Referenced to COMMON.
Program voltage source.
Referenced to V SOURCE LO.
Referenced to V SOURCE LO.
Program voltage source.
Referenced to V SOURCE LO.
Referenced to V SOURCE LO.
Referenced to V SOURCE LO.
Referenced to V SOURCE LO.
Referenced to V SOURCE LO.
Referenced to V SOURCE LO.

Hide quick links: