Table - Keithley 6517 Service Manual

Electrometer
Table of Contents

Advertisement

Table B-1
Calibration error responses (cont.)
Error number
Message
Charge Function Errors (cont.)
+394
"200nC ngain out of spec"
+395
"2uC pgain out of spec"
+396
"2uC ngain out of spec"
Temperature Function Errors
+399
"Temperature offset out of spec"
+400
"Temperature gain out of spec"
Humidity Function Errors
+409
"Hum. 50% offset out of spec"
+410
"Hum. 100% offset out of spec"
+411
"Hum. 50% gain out of spec"
+412
"Hum. 100% gain out of spec"
Offset Calibration Errors
+413
"Voltage Offset not converging"
+414
"Current Offset not converging"
Voltage Source Errors
+415
"VSRC 100V offset out of spec"
+416
"VSRC 100V pgain out of spec"
+417
"VSRC 100V ngain out of spec"
+418
"VSRC 1kV offset out of spec"
+419
"VSRC 1kV pgain out of spec"
+420
"VSRC 1kV ngain out of spec"
Factory Calibration Errors
+421
"Voltage Offset out of spec"
+422
"Current Offset out of spec"
+423
"Zero Check CAL Error"
Calibration Execution Errors
+424
"Date of calibration not set"
+425
"Next date of calibration not set"
+426
"Calibration not initialized"
+427
"Illegal Calibration Command"
Power-on Errors
+513
"Calibration data lost"
+514
"Calibration dates lost"
+515
"Calibration tolerances lost"
+516
"Calibration tables lost"
+517
"Voltage Offset lost"
+518
"Current Offset lost"
+519
"Installed option id lost"
+520
"Option card not supported"
+521
"Cal Card Data Error"
Note: Bus response to query includes error number, comma, and error message surrounded by double quotes.
Description
200nC range negative slope error
2µC range positive slope error
2µC range negative slope error
Temperature function offset error
Temperature function gain error
50% humidity offset error
100% humidity offset error
50% humidity gain error
100% humidity gain error
Voltage offset calibration error
Current offset calibration error
100V range offset error
100V range positive slope error
100V range negative slope error
1000V range offset error
1000V range positive slope error
1000V range negative slope error
Calibration date not set error
Calibration due date not set error
Calibration not initialized error
Illegal calibration command error
Calibration data lost error
Calibration dates lost error
Calibration tolerances lost error
Calibration tables lost error
Voltage offset value lost error
Current offset value lost error
Option ID lost error
Option card not supported error
Calibration unit data error
Calibration Messages