4 Start up and shut down | 4.1 Installing the MULTIflex coupling

4 Start up and shut down

Hazard from non-sterile products.
Infection hazard for dentist and patient.
▶ Before first use and after each use, prepare and sterilise the medical device
and accessories appropriately.
Dispose of the product in appropriate manner.
Infection hazard.
▶ Before disposal, reprocess and sterilise the product and accessories appro-
priately.
Damage from soiled and moist cooling air.
Contaminated and moist cooling air can cause malfunctions.
▶ Make sure that the supply of cooling air is dry, clean, and uncontaminated
according to EN ISO 7494-2.

4.1 Installing the MULTIflex coupling

Detachment of the medical device during treatment.
A medical device that is not properly locked can release from the MULTIflex
coupling during treatment.
▶ Before each use, check if the medical device is securely locked onto the
MULTIflex coupling.
▶ Screw the MULTIflex coupling onto the turbine hose.
▶ Open the water supply all the way using the spray ring on the MULTIflex
coupling.

4.2 Check the O-rings (MULTIflex)

Missing or damaged O-rings.
Malfunction and premature failure.
▶ Make sure that all O-rings are on the coupling and are undamaged.
Number of available O-rings: 5