Transmitting Data; Data Transmission Interval (Int 2) - wtw pH/ION 340i Operating Manual

Ph/ion meter
Hide thumbs Also See for pH/ION 340i:
Table of Contents

Advertisement

pH/ION 340i
Setting the
Data transmission
interval
4.7

Transmitting data

You have the following possibilities of transmitting data:
l One of the following options:
– With the AutoStore function (page 44), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the Data transmission interval function (Int 2), measured
values are periodically output on the interface (see below).
l With the Output data storage function (page 46), calibration data or
saved measured values are output on the interface.

4.7.1 Data transmission interval (Int 2)

The interval to the data transmission (Int 2) determines the chronolog-
ical interval between automatic data transmissions. After the selected
interval expires, the current data record is transmitted to the interface.
Note
The setting of the interval (Int 2) is only effective if the save interval
(AutoStore function) is switched off!
The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min, 5
min, 10 min, 15 min, 30 min, 60 min):
1
Press the <RUN/ENTER> key and hold it down.
2
Press the <RCL> key. Int 2 appears on the display.
ISE
UpH
S
I
n t
1
Time
Baud
8
O f
Day.Month
No.
Year
Ident
AutoCal DIN
Auto
Store
LoBat
AutoCal TEC
ConCal
3
Set the required interval between the saving procedures with
< > < >.
4
Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measur-
ing mode.
mol/L
2
mV/pH
% ppm
mg/L
°F
f
%
TP
°C
CalError
TempError
ARng
AR
RCL
Operation
53