Transmitting Data; Data Transmission Interval (Int 2) - wtw 2F30-100 Operating Manual

Ph/dissolved oxygen/conductivity measuring instrument
Table of Contents

Advertisement

Multi 340i
Setting the
Data transmission
interval
4.7

Transmitting data

You have the following possibilities of transmitting data:
l One of the following options:
– With the AutoStore function (page 44), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the Data transmission interval function (Int 2), measured
values are periodically output on the interface (see below).
– AutoStore (Int 1) covers the Data transmission interval
(Int 2).
l Press the <RUN/ENTER> key
This manually triggers a data transmission of the current measured
value to the serial interface at any time - independently of the select-
ed intervals.
l With the Output data storage function (page 46), calibration data or
saved measured values are output on the interface.

4.7.1 Data transmission interval (Int 2)

The interval to the data transmission (Int 2) determines the chronolog-
ical interval between automatic data transmissions. After the selected
interval expires, the current data record is transmitted to the interface.
The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min,
5 min, 10 min, 15 min, 30 min, 60 min):
1
Press the <RUN/ENTER> key and hold it down.
2
Press the <RCL> key. Int 2 appears on the display.
TDS
UpH
O
n
t 2
1
S
8
O
F
Baud
Time
Day.Month No.
Year
Ident
Tref25
Tref20
AutoCal DIN
LoBat
nLF
Lin
Cal
AutoCal TEC
3
Set the required interval between the saving procedures with
<▲> <▼>.
Sal
mV/pH
% mg/l
S/
m
cm
mbar
M
cm
1/
cm
Sal
F
°
C
/
%
K
TP
°
F
Auto
Store
ARng
AR
RCL
Operation
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

Multi 340i

Table of Contents