wtw pHotoFlex pH Operating Manual page 56

Led filter photometer with integrated ph function
Hide thumbs Also See for pHotoFlex pH:
Table of Contents

Advertisement

Operation
Sample of a record
Calibration evaluation
56
31.10.03
16:13
pHotoFlex Ser. no. 12345678
Calibration pH & ORP
Calibration date 31.10.03
Calibration interval
AutoCal TEC
Buffer 1
Buffer 2
Buffer 3
Voltage 1
Voltage 2
Voltage 3
Slope
Asymmetry
Sensor +++
After calibrating, the meter automatically evaluates the calibration. The
asymmetry and slope are evaluated separately. The worse evaluation
of both is taken into account. The evaluation appears on the display
and in the calibration record.
Sample dis-
Calibration
play
record
+++
++
+
-
Clean the electrode according
to the electrode operating
manual
----
----
Eliminate the error according to
chapter 6 W
HAT TO DO IF
16:13:33
7 d
4.01
7.00
10.01
184.0 mV
24.0 °C
3.0 mV
24.0 °C
-177.0 mV
24.0 °C
-60.2 mV/pH
4.0 mV
Asymmetry
[mV]
-15 ... +15
-20 ... +20
-25 ... +25
-30 ... +30
< -30 or
> 30
...
®
pHotoFlex
pH
Slope
[mV/pH]
-60.5 ... -58
-58 ... -57
-61 ... -60.5
or
-57 ... -56
-62 ... -61
or
-56 ... -50
... -62 or
... -50