pHotoFlex
pH
Display indication when
the measuring range is
undercut or exceeded
ba75975e05
10/2017
Display
"< [Lower limit of measuring range]"
instead of the measured value
"> [Upper limit of measuring range]"
instead of the measured value
Orientation value
Examples:
Photometry \ Concentration
0.16
< 0.20
83: A6/25 MC
16 mm
0.20 - 6.51 mg/l
01.02.05 15:12
Photometry \ Concentration
----
< 0.20
83: A6/25 MC
16 mm
0.20 - 6.51 mg/l
01.02.05 15:12
The orientation value can be quite inaccurate and should not be used
as a measure value!
Explanation
Measuring range undercut.
Remedy: Use a test with a lower
measuring range
Measuring range exceeded.
Remedy: Use a test with a high-
er measuring range or dilute the
sample
This value serves as a refer-
ence point for the selection of a
suitable test or dilution ratio.
If the measurement signal can-
not be evaluated (intensity too
high or too low), four bars ap-
pear ("----").
[BV]
Orientation value
mg/l
NH4-N
[BV]
Measurement signal cannot
be evaluated
mg/l
NH4-N
Operation
39