4.6.5 Calibration
pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is dis-
played. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the meter. Thus, you
should calibrate at regular intervals.
 After connecting another electrode
 When the sensor symbol flashes:
– after the calibration interval has expired
– after voltage interruption (e.g. empty batteries, empty
rechargeable battery)
You can use the buffer sets quoted in the table for an automatic calibra-
tion. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into account during
calibration.
Buffer set
WTW technical buffer solutions
NIST/DIN buffer solutions
The buffers are selected in the Configuration / pH & ORP / Cal. type
menu, see section 4.6.4).
pHotoFlex
Name on the
pH values at
display
25 °C
TEC
2.00
4.01
7.00
10.01
NIST/DIN
1.679
4.006
6.865
9.180
12.454