Prepare the 2606B for calibration adjustment
1. Connect the 2606B to the controller using an appropriate interface cable.
2. Turn on the 2606B and the test equipment and allow them to warm up for at least two hours
before performing calibration adjustment.
3. Make sure the communication interface parameters are set up properly (to configure the
interface, press the MENU key, and then select LAN; configuration of the USB interface is not
necessary).
Voltage calibration adjustment
1. Connect the 2606B SMU to the digital multimeter using the 4-wire connections shown in the
figure below.
2606B-901-01 Rev. B / May 2018
Figure 134: Connections for voltage verification and adjustment