Smux.contact.speed - Keithley 2606B System SourceMeter Reference Manual

Table of Contents

Advertisement

2606B System SourceMeter® Instrument Reference Manual
When the output is off and any of the following:
The output off mode is High-Z (error code 5048, "Contact check not valid with HIGH-Z OUTPUT
off")
The output off mode is Normal with the smuX.source.offfunc attribute set to
smuX.OUTPUT_DCVOLTS and the off current limit set to less than 1 mA (error code 5066,
"source.offlimiti too low for contact check")
The output off mode is Normal with the smuX.source.offfunc attribute set to
smuX.OUTPUT_DCAMPS and the source range is less than 1 mA (error code 5065, "I range too
low for contact check")
Example
if not smua.contact.check() then
smua.contact.speed = smua.CONTACT_SLOW
rhi, rlo = smua.contact.r()
print(rhi, rlo)
exit()
end
Also see
Contact check connections
Contact check measurements
smuX.contact.check()

smuX.contact.speed

smuX.contact.speed
This attribute stores the speed setting for contact check measurements.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
speedSetting = smuX.contact.speed
smuX.contact.speed = speedSetting
speedSetting
X
Details
This setting controls the aperture of measurements made for contact check. It does not affect the
smuX.measure.nplc aperture setting.
The speed setting can have a dramatic effect on the accuracy of the measurement (see
specifications).
2606B-901-01 Rev. B / May 2018
(on page 2-51)
(on page 2-50)
(on page 7-187)
(on page 7-189)
Affected by
Instrument reset
SMU reset
Recall setup
The speed setting. Set to one of the following:
0 or smuX.CONTACT_FAST
1 or smuX.CONTACT_MEDIUM
2 or smuX.CONTACT_SLOW
Source-measure unit (SMU) channel (for example, smua.contact.speed applies
to SMU channel A)
Section 7: TSP command reference
Check contacts against threshold.
Set speed for SMU channel A to slow.
Get resistance readings.
Output contact resistances to the host.
Terminate execution.
Where saved
Default value
Not saved
0 (smuX.CONTACT_FAST)
7-189

Advertisement

Table of Contents
loading

Table of Contents