The motor controller outlet is intended solely for connection of the motor controller and
must not be used for other devices, device components, or accessories.
We recommend DC-isolated cables for the connection of external equipment to the
XRCISE STRESS ECHO.
5.4.2

Cabeling

Make sure that nobody can step on or stumble across the power cord.
Do not place any objects on the cord as it might get damaged.
When transporting the table or changing its position, place all cables on top of the
table.
Set up the machine so that power switch and plug can easily be disconnected
5.5

Potential compensation

To avoid faults or in order to protect against contact voltages between touchable conductive
parts of the multifunction table and other devices in the patient environment, the
multifunction table should be connected to the clinic's internal potential equalization
system via a potential equalization cable. By default, the bed is equipped for potential
equalization. The grounding bolt is located beneath the control panel.
Never connect the potential compensation line to the water or gas supply line or any other
kind of pipe. Always use the appropriate potential compensation connection only.
Technical and visual modifications as well as misprints reserved
-