3.
Turn your device on using the mains switch. It is located on the frame at the foot
of the table. After your exercise machine has been connected to mains and
switched on, it automatically carries out an operating check. During this operating
check, you will be able to read the software version of the unit on the display.
Thereafter the main menu will appear
4.
Check if the display works. If this is not the case, make sure you followed the
steps above correctly. In addition, verify if there is electricity in the mains socket.
The training bike and the adjustment device control are connected separately. Thus, the
adjustment device operates independent from the training bike.
.4.1

Power supply

5
Use your exercise machine only with earthed (grounded) power sockets with 230 VAC /
50-60 Hz (see chapter A5). If you have any doubts about the power supply at the setup
location, ask your energy provider. Only use commercial 10 ampere automatic circuit
breakers (type B tripping characteristic). In the rare event that these automatic circuit
breakers should switch off when you switch on your machine, the circuit needs to be fused
with 10 A lead fuses or with a different type of tripping fuse (e.g. K-automat). In case of
doubt, ask your electrician.
Before connecting your cardiowise Gerätes exercise machine to your power supply system
compare the acceptable voltage and frequency on the name plate (next to power entry
module) with your local data.
Always connect your machine directly to the power outlet. If its possible do not use
extension cables or multioutlet power strips unless they are EN 60601-1 certified.
2 0
Technical and visual modifications as well as misprints reserved
Transport and Installation
Power Connection
-
© 2017 by ERGO-FIT GmbH & Co. KG