JDS Uniphase OLP-85 User Manual page 70

Power meters, smartclass fiber
Table of Contents

Advertisement

9 R
C
EMOTE
ONTROL
C
OMMANDS
Command string
:SENS:MEAS:MODE?
:SENS:POW:REF
<NR2>
:SENS:POW:REF?
:SENS:POW:UNIT
<NR1>
:SENS:POW:UNIT?
:SENS:THRS:PASS
:SENS:THRS:PASS?
:SENS:WLEN?
:SENS:WLEN:TBL
:SENS:WLEN:TBL?
:SENS:WLEN:TBL
:LMT
70
Parameter type / Response type / Unit / Info
Returns the active measurement mode.
Response type:
<NR1>
See :SENS:MEAS:MODE command for more information.
Sets the reference level in dBm.
Default:
0
Returns the reference level in dBm.
Response type:
<NR2>
See :SENS:POW:REF command for more information.
Select the unit for the absolute level.
Valid values:
0 (dBm)
1 (Watt)
Default:
0
Returns the unit for the absolute level.
Response type:
<NR1>
See :SENS:POW:UNIT command for more information.
Sets the threshold value "PASS" for the actual wavelength/unit dBm.
Default:
1)
0
Returns the threshold value "PASS" in dBm.
Response type:
<NR2>
See :SENS:THRS:PASS command for more information.
The actual "active" wavelength. This setup is driven by other setups and
results. (read only)
Table of selectable wavelength.
Returns the table of selectable wavelength.
See :SENS:WLEN:TBL command for more information.
Is a child map of :SENS:WLEN:TBL containing the threshold limit values
(unit dBm) for all selectable wavelength. Adding or deleting entries in
:SENS:WLEN:TBL will automatically update this map.
1)