4.3 Sensor Connections

Caution.
The connection terminals accept cables up to a maximum peripheral wire cross section of
2
2.5 mm
(12 SWG [10 AWG]).
Do not use a rigid conductor material as this can result in wire breaks.
Ensure the connecting cable is flexible.
To ensure the sensor cable length is sufficient, allow an additional 100 mm (4 in.) of cable to pass
through cable glands and into the housing.
Ensure the correct connections are made to suit the transmitter variant – ACA592–EC (see
Section 4.3.1, page 30).
OI/ACA592/EC–EN Rev. B
4 Electrical Installation
29