9.4 HART Common Practice Commands

Command
Description
No.
34
Write Damping
35
Write the PV Range
Write Configuration
38
Reset
40
Write Fixed Output Value
41
Start Self Test
42
Master Reset
44
Write PV Units
Write Trim Loop Current
45
Zero
Write Trim Loop Current
46
Gain
Write PV Transfer
47
function
Table 9.2 HART Common Practice Commands
78
Details
Enables the user to alter the PV damping value in seconds
Enables the low and high range values to be set for the
range.
Enables the configuration changed flag to be changed.
A value in the range of 3.8 to 21.5 can be written to fix the
output of the device.
A value of 0.0 should be written to fixed Output Value to
exit the device from fixed output mode.
Performs self-test function on the device.
Device can be reset.
Select the units in which the Primary Variable and Primary
Variable Range will be returned.
This will also select the units for Primary Variable Sensor
Limits and Primary Variable Minimum Span.
Enables user to trim the lower endpoint (for
example,~4.0mA) of the device's output.
Enables user to trim the upper endpoint (for example,
~20.0 mA) of the device's output.
User can alter the PV output linearity (used if a
function-generator is needed for output adjustment)
9 HART Communication
OI/ACA592/EC–EN Rev. B