ABB Endura ACA592 Operating Instructions Manual page 70

4-electrode conductivity transmitter
Hide thumbs Also See for Endura ACA592:
Table of Contents

Advertisement

Endura ACA592
4-electrode conductivity transmitter
Diagnostic Message
[Sub-error Message]
S072.040
Snsr temperature
over range
S074.001
Temperature
input error
S076.010
PV out of range limits
S078.004
PV out of physical limit
S082 039
Dirty sensor detected
Table 8.2 Diagnostic Messages (Continued)
68
Possible Cause(s)
[Related Condition]
Sensor temperature is above range.
[PROCESS]
Temperature related sensor board
electronics error.
[PROCESS]
The analog output for the Primary
Variable is beyond its scaling limit
and no longer represents the true
applied process.
[PROCESS]
The process applied to the sensor
for the Primary Variable is beyond
the operating limits of the device.
[PROCESS]
Sensor is dirty.
[SENSOR]
8 Troubleshooting and Diagnostics
Corrective Action
Verify sensor wiring.
Verify configuration related to temperature sensor
and temperature compensation.
(Maskable message)
Check temperature sensor connections.
If a temperature sensor is not connected, switch
temperature compensation to Manual mode.
If the problem persists, contact factory.
(Maskable message)
Adjust the working range if possible.
(Maskable message)
Verify sensor wiring.
Verify configuration setting for the sensor.
(Maskable message)
Verify sensor tip is clean.
Remove any foreign material.
Clean sensor.
Verify sensor responds to conductivity standard
solutions.
If sensor does not respond, electronically test
sensor.
Verify configuration.
OI/ACA592/EC–EN Rev. B