ABB Endura ACA592 Operating Instructions Manual page 85

4-electrode conductivity transmitter
Hide thumbs Also See for Endura ACA592:
Table of Contents

Advertisement

Endura ACA592
4-electrode conductivity transmitter
Equipment Markings
Intrinsic Safety – FM and CSA
FM
Class I, Div. 1, Groups A, B, C, D
Class II/III, Div. 1, Group E, F, G; T4 Ta = 60 ºC
CSA
Class I, Div. 1, Groups A, B, C, D
Class II, Div. 1, Groups E, F, G
Class III, Div. 1; T4
Intrinsic Safety – ATEX / IECEx
Approved for:
– II 1G Ex ia IIC T4
– II 1D Ex iaD A20 IP66 T135 °C, –20 °CT
Intrinsically Safe and Ex ia IIC Hazardous Area
Parameter
Maximum voltage
Maximum input
current
Maximum power
Internal inductance
Internal capacitance
Type n (Non-sparking) and Dust – ATEX / IECEx
Approved for:
II 3 G Ex nA IIC T4
Approve for:
II 3 D Ex tD A22 IP66 T135ºC, –20 °CT
Non-incendive – FM and CSA
(when installed in accordance with the installation drawing P0909 of Appendix B)
FM
Class I, Div. 2, Groups A, B, C, D
Class II/III, Div. 2, Group F, G; T4 Ta = 60 ºC
CSA
Class I, Div. 2, Groups A,B,C,D
Class II, Div. 2, Groups F, G
Class III, Div. 2; T4
Explosion-proof, Ignition-proof – FM and CSA
FM
XP, Class I, Div. 1, Groups A,B,C,D
Class II/III, Div. 1, Group E, F, G; T4 Ta = 60 ºC
CSA
Class I, Div. 1, Groups A,B,C,D
Class II, Div. 1, Groups E, F, G
Class III, Div. 1; T4
Flameproof and Dust Protection – ATEX / IECEx
Approved for:
II 2 G Ex d IIC T4
II 2 D Ex tD A21 IP66 T135ºC, –20 °CT
Agency Enclosure Ratings
IP66 and IP67
NEMA 4X
Approvals
CE Mark
The ACA592–EC including type B LCD display / configuration software meets all requirements for the CE mark in
accordance with the applicable directives 2004/108/EC (EMC), 2006/95/EC (LVD) and 94/9/EC (ATEX).
OI/ACA592/EC–EN Rev. B
60 °C
amb
Supply Circuit
U
= 30 V
i
I
= 160 mA
i
P
= 0,8 W
i
L
= 0,5 mH
i
C
= 5 nF
i
60 °C
amb
60 °C
amb
10 Specification
DS/ACA592-EN
83

Hide quick links:

Advertisement

Table of Contents
loading