Keysight Technologies X8712A User Manual page 31

Iot device battery life optimization solution
Table of Contents

Advertisement

Settings
Avg Channel 1
Sample & Duration
Sets the following criteria to
initialize and set up the software
to obtain real-time results.
Keysight X8712A User's Guide
Description
This waveform is digitized from Channel 1 to define the current event
mentioned in
page
44. This option is available in
the software acquired waveform from Channel 1 according to the zone
numbers selected.
Click Acquire once you change the zone number to reflect the
changes on the Avg Channel 1 waveform. Click Analyze will not
change the Avg channel 1 waveform.
Sample Size
Define the number of points in a measurement where the minimum
point is 1. The range is dependent on the number of channels below:
– One channel: up to 512K points.
– Two channels: up to 256K points.
– Three channels: up to 128K points.
– Four channels: up to 64K points
where K equals to 1024.
Time Interval (s)
Define the time period between samples, where the maximum is 0.1
seconds. The minimum time interval is dependent on the number of
channels below:
– One channel: 5.12 microseconds.
– One or two channels: 10.24 microseconds
– Three or four channels: 20.48 microseconds
Time Span (s)
Calculated by the software with a given Sample Size and Time Interval.
Getting Started
Show Legend
once
31