Micro-Disk Filter & Tubing Replacement - Philips EverFlo Q Service And Technical Reference Manual

Oxygen concentrator
Hide thumbs Also See for EverFlo Q:
Table of Contents

Advertisement

8.15 MICRO-DISK FILTER & TUBING REPLACEMENT
To remove the Micro-Disk Filter & Tubing
1. Remove the Filter Cover. Refer to the Filter Cover Replacement section for more information.
2. Remove the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
3. Remove the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
4. Remove the Flow Tubing from the DISS Outlet Fitting and the top connection of the Flow Meter.
5. Remove the Flow Tubing from the Micro-Disk Filter.
6. Remove the Pressure Tubing from the Pressure sensor located on the Main PCA (OPI Units Only).
7. Remove the other end of the Pressure Tubing from the bottom connection of the Flow Meter and
from the Pressure Regulator (OPI Units Only).
To install the Micro-Disk Filter & Tubing
1. Connect the Pressure Tubing to the Pressure Sensor located on the Main PCA (OPI Units Only).
2. Connect the Other ends of the Pressure Tubing to the Pressure Regulator and bottom connection
on the Flow Meter. Refer to Refer to Figure 8-26.
3. Connect the Flow Tubing as shown in Refer to Figure 8-27.
Page 79
Figure 8-28: PRESSURE TUBING CONNECTIONS
Figure 8-29: FLOW TUBING INSTALLATION
LO TM
E
F
S
& T
VER
ERVICE
I
ECHNICAL
NFORMATION
1039055, VER. 06