2. Install the replacement blue tube and attach the clamp with the hose clamp pliers. See assembled kit
below.
3. Slide the protective plastic pipe over the blue tube all the way down to the hose clamp.
4. Attach the cable tie above the plastic pipe and trim the excess cable tie.
5. Install the foam in the compressor section of cabinet if the foam was not already installed as original
equipment.
To replace the Philips Respironics compressor's inlet tube assembly
The inlet tube assembly removal and replacement procedure detailed in this Procedure section is only
applicable for Philips Respironics manufactured compressors.
Page 73
Figure 8-27: Assembled Kit
NOTE