7.4

SYSTEM PRESSURE TEST TABLE

SYMPTOM
System Pressure Test
Abnormal reading
Fluctuating
pressure
Low pressure
reading
Page 39
CAUSE
High-pressure
reading above 25
psig @ 60 Hz @ 5
LPM flow.
Difference in Peak
pressure is > 1 psig.
Low pressure reading
below 20 psig @ 5
LPM
LO TM
E
F
S
& T
VER
ERVICE
VERIFICATION
Check oxygen
percentage at 5
LPM. If Low?
Check oxygen
percentage at 5
LPM. If Low?
Check oxygen
percentage at 5
LPM. If Low?
Check for leaks
Check tubing
Check fittings
Check for dirty filters
Check Compressor
for System Pressure
I
ECHNICAL
NFORMATION
CORRECTIVE
ACTION
Check for proper
valve cycling
Replace Sieve
Canister
Check for leaks
Replace Sieve
Canister
Replace tubing
Replace fittings
Replace filters
Replace
Compressor
1039055, VER. 06