Avoid handling the molecular sieve material. Respironics recommends the return of the sieve
canister assembly to Respironics for any service that involves sieve disposal.
Use extreme caution when handling the compressor capacitor as it holds an electrical charge
until is it properly discharged.
Device operation above or outside of the voltage, LPM, temperature, humidity and/or altitude
values specified may decrease oxygen concentration levels.
If the EverFlo Oxygen Concentrator has been subjected to sub-freezing temperatures for an
extended period, it should be allowed to warm up to the stated operating temperatures before
power up. Failure to do so could result in improper performance and or alarm conditions until
the device reaches normal operating temperatures.
2.2

CAUTIONS

U.S. federal law restricts this device to sale by or on the order of a physician.
Do not place liquids on or near the device.
If liquid is spilled on the device, turn the power off and unplug from electrical outlet before attempting to
clean up a spill.
When using liquid leak detector, be careful not to allow it to contact electrical parts.
Make sure connections of fittings, tubing, and hoses are secure.
Be cautious when using thread sealants because they can cause extensive damage to the internal parts
of the device if allowed within tubing or fittings.
Page 7
WARNINGS
CAUTIONS
LO TM
E
F
S
& T
VER
ERVICE
I
ECHNICAL
NFORMATION
1039055, VER. 06