Philips EverFlo Q Service And Technical Reference Manual page 35

Oxygen concentrator
Hide thumbs Also See for EverFlo Q:
Table of Contents

Advertisement

SYMPTOM
LED failure
Unit is inoperable
Start-up alarms in-
operable
Page 35
CAUSE
No power to LED's
Main PCA failure
Piezoelectric speaker
failure
LO TM
E
F
S
& T
VER
ERVICE
VERIFICATION
Verify the power cord
is plugged in
Verify power in wall
outlet
Verify that all wiring
harnesses are
connected
Verify the Main PCA
is not damaged
Verify the wiring
harnesses are
connected
Verify the wiring
harness is not
damaged
Verify the condition
of the fuse on PCA
Verify the condition
of the Power Cord
Verify the solder
joints on the
piezoelectric speaker
I
ECHNICAL
NFORMATION
CORRECTIVE
ACTION
Connect the
power cord to AC
power source
Check the
household fuse or
circuit breaker
Connect all wiring
harnesses
Replace the Main
PCA
Connect the wiring
harnesses
Check the wiring
harnesses for
continuity
Test the fuse for
electrical
continuity
Verify the Power
Cord continuity
Replace Power
Cord
Replace the Main
PCA
Replace the Main
PCA
1039055, VER. 06