Everflo And Everflo Q Pressure Regulator Assembly Replacement - Philips EverFlo Q Service And Technical Reference Manual

Oxygen concentrator
Hide thumbs Also See for EverFlo Q:
Table of Contents

Advertisement

8.7
EVERFLO AND EVERFLO Q PRESSURE REGULATOR ASSEMBLY
REPLACEMENT
To remove the Pressure Regulator Assembly
1. Remove the Filter Cover. Refer to the Filter Cover Replacement section for more information.
2. Remove the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
3. Remove the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
4. Gently remove the yellow pressure line coming from the Pressure Regulator Adaptor.
Figure 8-9: REMOVE YELLOW PRESSURE LINE
5. Unlock the Pressure Regulator Adaptor by twisting upwards until the Pressure Regulator stops.
6. Lift the Pressure Regulator Assembly straight up until it clears the Sieve Canister Assembly.
7. Using pliers, loosen the clamp and disconnect the flow tubing from the Pressure Regulator
Assembly.
To install the Pressure Regulator Assembly
1. Using pliers, clamp the flow tubing to the Pressure Regulator Assembly.
2. Place the Pressure Regulator Assembly in place by lining the Pressure Regulator Adaptor up with
the grooves in the Sieve Canister Assembly.
3. Lock the Pressure Regulator Assembly in place by turning the Assembly downwards.