Philips EverFlo Q Service And Technical Reference Manual page 33

Oxygen concentrator
Hide thumbs Also See for EverFlo Q:
Table of Contents

Advertisement

SYMPTOM
High pressure
Constant Audible
Alarm (no LED's
on)
Page 33
CAUSE
Pressure relief valve
Valve Solenoid failure
Fuse failure
LO TM
E
F
S
VER
VERIFICATION
Verify that tubing is
not kinked or
pinched
Verify that the tubing
is not damaged
Verify the valve
wiring harness is
connected at the
Main PCA
Verify the
Compressor is
working correctly
Verify the Valve
Solenoid is working
properly
Verify the valve
wiring harness is
connected at the
Main PCA
Verify the continuity
of the fuse
& T
I
ERVICE
ECHNICAL
NFORMATION
CORRECTIVE
ACTION
Reposition the
tubing
Replace the
damaged tubing
Replace Valve
Solenoid
Assembly
Replace the
Compressor
Replace the Valve
Body
Reconnect the
wiring harnesses
and retest
If problem
continues, install
old valve body
and new pilot
valves
Replace the fuse
1039055, VER. 06