Philips EverFlo Q Service And Technical Reference Manual page 106

Oxygen concentrator
Hide thumbs Also See for EverFlo Q:
Table of Contents

Advertisement

9. Verify the MSP software version is the most current. If not the most current version, then continue
with the Upgrade process.
10. Select the File and then the Upgrade Firmware option.
If the Upgrade Firmware option is unavailable, the Manifest file has not been properly loaded.
11. If the PCA is original to the unit, the model number and serial number will match the device label
and show up in the By Identification box. See Note below. If not, then proceed to step:18
The technician must determine that the device is completely original and has not been serviced previously. If
the unit was serviced in the past then upgrade the firmware By Feature Only. PCAs equipped with LCD hour
meters require upgrading via the By Feature method.
12. Upgrade when a single known configuration shows in the bottom box.
13. Click the Upgrade button and the EverFlo will alarm as the firmware is installed.
14. Click the Print button to print an Upgrade Record form. Print Function: see with an example of the
Upgrade Record. The technician is required to fill out the information on the form and keep it with
the Service paperwork (Only Chance to Do This)
15. Click the OK button to return to the main screen.
Page 106
NOTE
NOTE
LO TM
E
F
S
& T
VER
ERVICE
I
ECHNICAL
NFORMATION
1039055, VER. 06