2. Slide the inlet boot bellows (see Section 8.2) onto the tee. Secure with the included cable tie. Cut any
excess cable tie and discard.
3. Insert the inlet tube assembly into the compressor housing making sure to orient the inlet boot bellows
on the same side of the compressor as the valve cover barb.
4. Secure the inlet tube assembly to the compressor housing by wrapping the two cable ties in the inlet
tube assembly repair kit around the elbow and tee and through the compressor housing cutouts. Cut any
excess cable tie using side cutters.