8.5
REAR CABINET/POWER CORD REPLACEMENT
To remove the Rear Cabinet/Power Cord
1. Place the concentrator face down on an ESD Protected surface.
2. Remove the Filter Cover. Refer to the Filter Cover Replacement section for more information.
3. Remove the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
4. Remove the two phillips screws from the handle of the Rear Cabinet.
5. Remove the four hex screws from the Rear Cabinet.
6. Lifting the Rear Cabinet slightly up from the Front Cabinet, fold the Rear Cabinet away from the
Front Cabinet.
7. Cut the two tie wraps securing the Power Cord to the Sieve Canister.
Page 46
Figure 8-3: SCREW REMOVAL/INSTALLATION
Figure 8-4: EVERFLO REPAIR POSITION
LO TM
E
F
S
& T
VER
ERVICE
I
ECHNICAL
NFORMATION
1039055, VER. 06