Philips EverFlo Q Service And Technical Reference Manual page 36

Oxygen concentrator
Hide thumbs Also See for EverFlo Q:
Table of Contents

Advertisement

SYMPTOM
Low oxygen output
Low Oxygen
percentage
indicator not
functioning
Continuous red
LED
Warm cabinet
temperature
Page 36
CAUSE
Sieve bed failure
Compressor failure
Main PCA failure
Cooling Fan failure
LO TM
E
F
S
VER
VERIFICATION
Verify the oxygen
output with a
calibrated oxygen
analyzer
Verify the
Compressor is
working properly
Verify that there are
no leaks
Verify Compressor is
operating properly
Verify the OPI output
Verify the Cooling
Fan connectors are
seated properly
Verify there is
continuity on
electrical fan wiring
harness
Verify there is power
to the Cooling Fan
Verify the Cooling
Fan is not obstructed
& T
I
ERVICE
ECHNICAL
NFORMATION
CORRECTIVE
ACTION
Replace the Sieve
Canister
Replace the
Compressor
Replace tubing
and clamps
Replace
Compressor
Replace the Main
PCA
Reseat the
connectors
Replace the fan
assembly
Reseat the
connectors
Remove the
obstructions from
the Cooling Fan
1039055, VER. 06