To Remove The Multimode Source - Agilent Technologies G1978B User Manual

Multimode source for 6410 triple quad lc/ms
Hide thumbs Also See for G1978B:
Table of Contents

Advertisement

To remove the multimode source

Do the following steps to remove the multimode source.
1 Turn off the multimode source temperatures and flows:
Do not touch the multimode source or the capillary cap. They may be very hot. Let
WA R N I N G
the parts cool before you handle them.
Never touch the source surfaces, especially when you analyze toxic substances or
WA R N I N G
when you use toxic solvents. The source has several sharp pieces which can pierce
your skin including the APCI corona needle, vaporizer sensor and counter current
electrode.
Do not insert fingers or tools through the openings on the multimode chamber. When
WA R N I N G
in use, the capillary and capillary cap are at high voltages up to 4 kV.
2 Wait approximately 20 minutes or until the source is cool.
3 Open the CDS door at the front of the MS to access the cables.
4 Disconnect the ESI high voltage charging electrode cable.
5 Disconnect the APCI Needle Interlock, and multimode HV cable.
6 Unscrew the nebulizer gas line from the nebulizer.
7 Unscrew the LC sample tubing from the nebulizer.
8 Open the latch on the source and open the source.
9 Remove the multimode source from the spray chamber mount.
10 Place the source shipping cover on the source.
Multimode Source for 6410 Triple Quad LC/MS User Guide
a Change the Context view to Acquisition.
b Click the MS QQQ tab.
c Turn off all voltages and temperatures in the Source tab.
d Wait approximately 20 minutes for the source to cool down.
Installation

To remove the multimode source

2
23

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents