2
Installation
To convert from multimode to ESI or APCI
To convert from multimode to ESI or APCI
Never touch the source surfaces, especially when you analyze toxic substances or
WA R N I N G
when you use toxic solvents. The source has several sharp pieces which can pierce
your skin including the APCI corona needle, vaporizer sensor and counter current
electrode.
1 Unscrew and remove the multimode spray shield with the field shaping
2 Install the new source and the standard spray shield, making sure that the
3 For an APCI ion source, connect the vaporizer heater cable and the APCI
4 For all sources, reconnect the nebulizer gas line tubing and the LC/MS
24
electrodes.
hole in the spray shield is in the 12 o'clock position.
high voltage cable.
sample tubing.
Multimode Source for 6410 Triple Quad LC/MS User Guide