To convert from ESI or APCI to the multimode source
If you are installing this source on this instrument for the first time, follow the steps in
C A U T I O N
"Installation"
1 Turn off the multimode source temperatures and flows:
2 Wait for the source to cool (until temperatures are at least below 100°C).
3 Disconnect the nebulizer gas tubing from the currently installed ion source.
4 Disconnect the LC/MS sample inlet tubing.
5 If the APCI source is installed, remove the APCI vaporizer heater cable and
6 Remove the currently installed ion source.
7 Unscrew and remove the spray shield. See
Do not touch the multimode source or the capillary cap. They may be very hot. Let the
WA R N I N G
parts cool before you handle them.
Do not insert fingers or tools through the openings on the multimode chamber. When
WA R N I N G
in use, the capillary and capillary cap are at high voltages up to 4 kV.
Multimode Source for 6410 Triple Quad LC/MS User Guide
on page 17.
a Change the Context view to Acquisition.
b Click the MS QQQ tab.
c Turn off all voltages and temperatures in the Source tab.
d Wait approximately 20 minutes for the source to cool down.
APCI high voltage cable.
To convert from ESI or APCI to the multimode source
Figure
Installation
13.