Step 1. Prepare To Install - Agilent Technologies G1978B User Manual

Multimode source for 6410 triple quad lc/ms
Hide thumbs Also See for G1978B:
Table of Contents

Advertisement

2
Installation

Step 1. Prepare to install

Step 1. Prepare to install

The Multimode Enablement Kit, G1978-60451, is shipped with the multimode
source. This kit needs to be installed before the multimode source is used.
Note that the multimode source and its accessories are to be installed by an
Agilent Customer Engineer.
1 Check that the Multimode Enablement Kit contains the following parts:
Figure 3
From left to right: G1960-60858, G1960-61015 and G1960-60873
2 Install the APCI Enablement Kit, G1947-60451, which is shipped with the
Figure 4
From left to right: G1946-80058, G1960-60802 and G1960-60856
18
Multimode Bd HV Cable, p/n G1960-60858
Multimode HV PCA, p/n G1960-61015
Multimode Bd Power/Data Cable, p/n G1960-60873
multimode source.
The APCI Enablement kit contains the following parts:
Fast APCI HV Supply, p/n G1946-80058
Valve BD-APCI Supply Cable, p/nG1960-60802
Valve BD-APCI Needle Interlock Cable, p/n G1960-60856
Multimode Source for 6410 Triple Quad LC/MS User Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents