Reference
G2432A MM-APCI Calibration Ions
Fragment ion signals are progressively stronger at higher m/z by increasing CID energy.
N O T E
G2432A MM-APCI Calibration Ions
Purine+H
HP-0321+H
HP-0622+H
HP-0921+H
HP-1521+H
HP-2121+H
Fragment ion signals are progressively stronger at higher m/z by increasing CID energy.
N O T E
54
fragment
fragment
Purine-H
fragment
fragment
fragment
fragment
Positive Ion (m/z)
121.051421
322.048699
622.029499
922.010300
1307.969049
1521.971900
1807.937049
2121.933500
Negative Ion (m/z)
119.035771
556.001426
805.985476
1305.953400
1805.921399
Multimode Source for 6410 Triple Quad LC/MS User Guide
Empirical Formula
C5.H5.N4
C6.H19.O6.N3.P3
C12.H19.O6.N3.P3.F12
C18.H19.O6.N3.P3.F24
C25.H17.O6.N3.P3.F40
C30.H19.O6.N3.P3.F48
C35.H17.O6.N3.P3.F60
C42.H19.O6.N3.P3.F72
Empirical Formula
C5.H3.N4
C10.H15.O6.N3.P3.F10
C15.H15.O6.N3.P3.F20
C25.H15.O6.N3.P3.F40
C35.H15.O6.N3.P3.F60