Total Reset To Factory Defaults; Battery Replacement - Check-line DC**-3000PS series Operating Instructions Manual

Coating thickness gauge
Table of Contents

Advertisement

To print statistical values:
1. Press DOWN repeatedly until Statistics appears on display.
2. Press OK. Printout Of Statistics appears.
3. Point instrument with the IR transmitter (top end face) towards the IR adapter
installed on the printer. The distance should be approx. 12" to 20" (30cm to
50cm).
4. Press OK. The data are transmitted, and Printing... will briefly appear, followed
by the start screen with the four dashes - - - -.
To continue measuring, press ESC repeatedly until the start screen with the four
dashes - - - - is displayed again.
6.4
Deletion of statistical values and of single values
1. Press DOWN repeatedly until Statistics appears on display.
2. Press OK. Printout Of Statistics appears.
3. Press DOWN repeatedly until Deletion Of Statistics appears.
4. Press OK. The statistical values with the single values are deleted. Statistics
deleted will briefly appear on the display, then the start screen with the four
dashes - - - -.
Quick deletion of statistics and stored readings
1. Press ESC repeatedly until the start screen with the four dashes - - - - is again
displayed.
2. Press ESC until Delete Statistic? appears on the display.
3. Press ESC again. Statistics Deleted will briefly appear, then the start screen
with the four dashes - - - -.
© Checkline-Europe
-16-
3.5
Manually Setting Measurement Mode (DCFN only)
In some cases, especially with varnish applied to zinc on steel, it is advisable to
manually set the measuring mode—F mode (Ferrous) for measurements on
steel/iron, N mode (Non-Ferrous) for measurements on non-ferrous metals.
1. Press DOWN repeatedly until display shows Options.
2. Press OK. Measuring Unit appears on the display.
3. Press DOWN repeatedly until Measuring Mode appears.
4. Press OK. Auto FN Identifications appears.
5. Press DOWN repeatedly until either Ferrous (F) or Non-Ferrous (N) or Auto
FN appears on the display.
6. Press OK to confirm the measurement mode that you want to activate. The
display shows - - - - and either Auto FN or Ferr or Non-Ferr. See diagram of
dsiplay on page 5.
3.6

Total Reset To Factory Defaults

This procedure deletes all single and statistical values. It also deletes set calibration
values and optional items and resets the gauge to the factory default.
1. Switch off the gauge
2. Press and hold the ESC key (item E, page 3). At the same time turn the power
back on.
3. An acoustic signal will sound and all factory default settings are restored.
4. Release both keys and the gauge will restart with all factory default settings
restored.
3.7

Battery Replacement

3000PS gauges use two AA alkaline or
rechargeable batteries. When battery power
becomes low the Battery Indicator will flash.
The gauge will continue to operate, but batteries
should be replaced as soon as possible When
battery power is depleted, the Battery Indicator
remains on continuously, indicating that less
than 60 seconds of power remain shut down.