 Confirm all flashing indicated parameters one by one by
them as necessary.
 Open heating hood and place removal aid with an empty sample dish on the
sample dish retainer. Make sure that the sample dish is resting flat on the
sample dish retainer. Always work with a removal aid as this will enable safe
working and prevent burns.
 Close heating hood and tare with
 Open heating hood.
Distribute sample evenly on the sample tray. A practical sample amount is
typically approx. 3- 5 g.
 Close heating hood.
 Start drying process by
The status display
You can watch the drying process on the display screen.
The current temperature, lapsed time as well as the current interim result will
be continuously updated and displayed.
The icons of the enabled heating profile and the switch-off criterion are
flashing.
Using
the display con be switched over to the different displays of
results.