6 Transport and storage

6.1 Testing upon acceptance

When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.

6.2 Packaging / return transport

 Keep all parts of the original packaging for a possibly required
return.
 Only use original packaging for returning.
 Prior to dispatch disconnect all cables and remove
loose/mobile parts.
 Secure all parts against shifting and damage.

7 Unpacking, Setup and Commissioning

7.1 Installation Site, Location of Use

The instrument is designed to achieve reliable weighing results under normal
conditions of use.
You will work accurately and fast, if you select the right location for your moisture
analyser.
On the installation site observe the following:
!
Remove explosion prone and easily flammable material in the immediate
vicinity.
Emerging vapours, sample dish and all parts of the sample chamber are hot!
!
Protect the instrument against direct draughts due to open windows and
doors.
!
Avoid extreme heat and temperature fluctuations e.g. due to installation next
to radiators.
!
Do not expose the instrument to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the instrument)
may occur if a cold instrument is taken to a considerably warmer environment.
In this case, acclimatize the disconnected instrument for ca. 2 hours at room
temperature.
!
Avoid direct sunlight
!
The air humidity should be between 45% and 75%, non-condensing.