2
Brief instructions
2.1
Mechanical connection
2.1.1
Connection flanges
2.1.2
Inlet / outlet piping
Operating mode
Unidirectional operation
Unidirectional operation
Bidirectional operation
Bidirectional operation
1 Depending on the nominal width.
th
September, 4
2018
This chapter does not replace the rest of the operating instruc-
tions. It shows only a brief section of the steps necessary in order
to make the device ready for operation.
The chapter is only directed at experienced users.
•
Observe the chapter safety.
Section 5, "Safety" on page 37
Detailed information for this content can be found under:
Section 7, "Construction and Planning" on page 65
Section 8, "Installation" on page 77
Section 9, "Start Up" on page 115
Section 12.4, "Troubleshooting" on page 157
1 Make sure that the device and the connection flange have
the same pressure rating / flange standards.
2 Make sure that the device is sealed with the appropriate
seals.
Inlet piping
10 D (no flow conditioner)
3 / 5 D (with RMG or stan-
1
dardized flow conditioner)
10 D (no flow conditioner)
3 / 5 D (with RMG or stan-
1
dardized flow conditioner)
– See also "Inner diameter of connecting spool pieces" on
Ultrasonic Flowmeter USM GT400
Outlet piping
3 D
3 D
10 D (no flow conditioner)
3 / 5 D (with RMG or stan-
dardized flow conditioner)
page 168
2 Brief instructions
Temperature sensor
position
1.5 D to 5 D
1.5 D to 5 D
3 D to 5 D
1
2 D to 5 D
1
5